🎛️Meter Conversion & Electrical Measurements - shunt, multiplier, ohmmeter, EMF
The meter-related parts of school electrical practicals in one place. Enter a galvanometer’s Ig and Rg to get the shunt for an ammeter or the series multiplier for a voltmeter, watch the needle swing on a dual scale, and see how a resistor error biases the reading. The ohmmeter page draws the non-linear scale (0 on the right, ∞ on the left) and explains the mid-scale resistance and why an aged cell reads high. The voltmeter–ammeter page compares internal and external connections against √(R_A·R_V). The EMF page fits E and r from noisy simulated U–I data for both circuits, and the last page covers the Wheatstone bridge and the half-deflection method.
Free · No sign-up · Runs in your browserLast updated:
Loading tool...
How to use Meter Conversion & Electrical Measurements
- 1On Meter conversion enter Ig, Rg and the target range to get the resistor
- 2On Ohmmeter change E and Rx, watch the non-linear scale and try an aged cell
- 3On the voltmeter–ammeter and EMF pages switch the connection and compare with true values
- 4On Bridge & half-deflection balance the bridge and see how the half-deflection error depends on E
FAQ
- How do I convert a galvanometer into an ammeter or a voltmeter?
- At full scale the coil has U_g = IgRg. For an ammeter of range I, a parallel shunt must carry I − Ig: R_sh = IgRg / (I − Ig) = Rg / (n − 1) with n = I/Ig. For a voltmeter of range U, add a series multiplier R_m = U/Ig − Rg = (n − 1)Rg with n = U/U_g. With Ig = 1 mA and Rg = 100 Ω, a 0.6 A ammeter needs about 0.167 Ω and a 3 V voltmeter needs 2900 Ω.
- When should the ammeter go inside or outside the voltmeter?
- Ammeter inside: the voltmeter also reads the ammeter’s drop, so R_m = R + R_A, too high. Ammeter outside: the ammeter also carries the voltmeter current, so R_m = R·R_V/(R + R_V), too low. Compare R with √(R_A·R_V): use the internal connection for larger R and the external connection for smaller R.
- Why is an ohmmeter scale uneven, and what is the mid-scale resistance?
- After zeroing, the meter’s internal resistance is R_in = E/Ig. With Rx connected, I/Ig = R_in/(R_in + Rx), which is not proportional to Rx, so the scale runs from 0 on the right to ∞ on the left and crowds to the left. When Rx = R_in the needle sits exactly in the middle, so the mid-scale value equals the internal resistance. Readings are most accurate near the middle.
- What errors do the two EMF circuits give?
- With the voltmeter straight across the cell (the usual textbook circuit), the voltmeter draws current and the equivalent source is the cell in parallel with R_V: E_m = E·R_V/(R_V + r) and r_m = r·R_V/(R_V + r), both slightly low but tiny because R_V ≫ r. With the ammeter in series with the cell, E_m = E exactly but r_m = r + R_A, which for a dry cell can exceed r itself.
- Why does the half-deflection method read the coil resistance low?
- You set full scale with S₂ open, then close S₂ and adjust the decade box R₂ to half scale and take R₂ as Rg. Closing S₂ lowers the circuit resistance, so the total current rises a little above Ig and R₂ carries more than Ig/2, hence R₂ < Rg. Exactly R₂ = A·Rg/(A + Rg) with A = R₁ + r, an error of about −Rg/A. A larger supply EMF needs a larger R₁ and gives a smaller error. The voltmeter half-deflection method reads high instead.
Features
- Ammeter and voltmeter conversion: shunt, multiplier, range factor, new meter resistance
- Damped swinging needle on a dual scale, with the reading error from an off-value resistor
- Ohmmeter non-linear scale, mid-scale resistance and aged-cell error
- Internal vs external connection: error formulas, critical resistance and error curves
- EMF and internal resistance: simulated data, U–I fit and the theoretical bias of both circuits
- Wheatstone bridge galvanometer current, current and voltage half-deflection errors